605xA and 606xB: Electronic Load Power-On State
Question: When powered on, what is the wake-up voltage and current setting of the 606xB and the modules
installed in the Agilent 605xB Series electronic loads?
Answer: The 606xB and the modules installed in the 605xA mainframe have 7 user-definable recall states which
allow saving the settings of the load for later recall. One of these states, location 0 is non-volatile; this is called the
wake-up state. The 606xB and the modules installed in the 605xA will be set to the state stored in location 0 at
power turn-on. If no state has been saved to location 0, then it will still contain the factory-default state. The factory
default state normally would be 0 V and 0 A. See *SAV in Programming Manual 06060-90005.