Specifications
605xA and 606xB: Electronic Loads in Series/Parallel
Question: Can I use Agilent Electronic Loads in series and in parallel?
Answer: Agilent electronic loads ARE designed to be operated in parallel for more current, but NOT in series for
more voltage. Loads are fully protected against damage from current overloads, but will be damaged by voltage
above the maximum voltage rating.