Specifications

Constant Current Operating Mode
Question: How do I put the power supply in the constant current mode?
Answer: The power supply cannot be "put" into the constant current mode. The output settings of the power supply
combined with the ohmic value of the particular load determine whether or not the power supply is in constant
current.
For example: The power supply inherently resides in the constant voltage mode. If the output voltage were set to 24
volts and a 6 load were placed across the output terminals, Ohm's Law would require that 4 amps would flow
(24V/6). This presumes that the constant current setting of the power supply were set to a value greater than 4
amps; lets say 5 amps. Now, if the 6 load were replaced by a 2 load, Ohm's Law would suggest that 12 amps
(24V/2 ) would flow. However, the power supply is set to go into constant current at 5 amps. Therefore, the actual
output voltage would be 10 volts (2 x 5A). The power supply will now remain in constant current for values of
load = 0 R < 4.8? . Once the ohmic value of the load becomes greater than 4.8 (24V/5A), the power supply will
again revert to constant voltage operation at the value of 24 volts.