OPERATING MANUAL 603xA FAMILY AUTORANGING SYSTEM DC POWER SUPPLIES 6030A, 6031A, 6032A, 6035A, 6033A, 60380A 6010A, 6011A, 6012B, 6015A, 6023A, 6028A
Standard Commands for Programmable Instruments (SCPI)
95
*TRG
Meaning and Type
Trigger
Device Trigger
Description
This command, which is essentially the same as the Group Execute Trigger (<GET>) and the subsystem TRIG[:IMM]
signals, generates a trigger to the power supply.
Command Syntax 
*TRG
*TST?
Meaning and Type
Test 
Device Test
Description
This query causes the power supply to do a self test and report any errors.
Query Syntax  *TST?
Returned Paramters  <NR1>
0 Indicates power supply passed self-test.
Nonzero  Indicates a self-test failure.
*WAI
Meaning and Type
Wait to Continue 
Device Status
Description
This command instructs the power supply not to process any further commands until all pending operations are completed.
"Pending operations" are as defined under the 
*OPC command. *WAI can be aborted only by sending the power supply a
GP-IB 
DCL (Device Clear) command.
Command Syntax  *WAI
ABOR
This command cancels any pending triggered levels stored by the CURR:TRIG or VOLT:TRIG commands. The pending
level is set equal to the corresponding immediate value. 
ABORt also resets the WTG bit in the Operation Condition status
register (see "Status Reporting"). If 
INIT:CONT ON has been programmed, the trigger system rearms itself immediately
after 
ABORt, thereby setting WTG. ABORt is executed at power turn on.
Command Syntax  ABORt
Examples ABOR ABORT