Table 2-1. Test Equipment Required
TYPE REQUIRED CHARACTERISTICS USE RECOMMENDED MODEL
Oscilloscope Sensitivity: 1 mV
Bandwidth: 20MHz & 100MHz
Input: differential, 50 & 10M
P,T Agilent 54504A
RMS Voltmeter True rms, 10MHz bandwidth
Sensitivity: 1 mV Accuracy: 5%
P Agilent 3400A
Logic Pulser 4.5 to 5.5Vdc @ 35mA T Agilent 546A
Multimeter Resolution: 100nV
Accuracy: 0.0035%, 6½ digit
P,C,T Agilent 3458A
Signature Analyzer -- T Agilent 5004A
GPIB Controller Full GPIB capabilities C,T,P HP Series, 200/300
Current Probe No saturation at 30Adc
Bandwidth: 20Hz to 20MHz
P Tektronix P6303 Probe/
AM503 Amp/
TM500 Power Module
Electronic Load Voltage range: 60Vdc P,C Agilent 6060A
Current range: 30Adc
Power range: 250W
Open and short switched
Power Resistor*
Value: 0.25 >200W (6033A)
P,C
Value: 2.3 >200W (6038A)
Current-Monitoring
Shunts
Value: 10m ± 0.04% @ 100W (6033A)
PC: 0.0004%/W
P,C** Guildline 9230/100
Value: 100m ± 0.04% @ 25W (6038A)
PC: 0.0004%/W
Guildline 9230/15
Calibration and Test
Resistors
Value: 100, 5%, 1W
1, 5%, ½W
1K, 5%, ¼W
2K, 0.01%, ¼W
C,T
Terminating
Resistors (2)
Value: 50 ±5%, noninductive
P
Blocking
Capacitors (2)
Value: 0.01µF, 100Vdc
P
Common-mode
Toroidal Core
3.7µH/turn
2
23mm I.D.
P Ferrox-Cube
500T600-3C8,
Agilent 9170-0061
Switch* SPST, 30A @60V P
DC Power Supply Voltage range: 0-60Vdc
Current range: 0-3Adc
C,T Agilent 6024A
P = performance testing C = calibration adjustments T = troubleshooting
* Not required if using electronic load.
** Less accurate, and less expensive, current-monitor resistors can be used, but the accuracy to which current programming
and readback can be checked must be reduced accordingly.