Service manual

112
Performance Test Record - Agilent 6038A POWER SUPPLY (Page 2 of 2)
MODEL Agilent 6038A
Report No.______________ Date_____________________
Page
No.
Test Description
Minimum Spec. Results
*
Maximum
Spec.
Measurement
Uncertainty
Constant Voltage Tests
21 Voltage Programming
and Readback
Low Voltage (0.090V)
Front Panel Display
GPIB Readback
0.050
V
OUT
- 0.050
V
OUT
- 0.050
________V
________V
________V
0.130
V
OUT
+ 0.050
V
OUT
+ 0.050
3
µV
3
µV
3
µV
High Voltage (60.0V)
Front Panel Display
GPIB Readback
59.939
V
OUT
- 0.092
V
OUT
- 0.092
________V
________V
________V
60.061
V
OUT
+ 0.092
V
OUT
+ 0.092
845
µV
845
µV
845
µV
22 Load Effect (20.00V)
V
OUT
- 0.005 _______mV V
OUT
+ 0.005 20µV
22 Source Effect (60.0V)
V
OUT
- 0.008 _______mV V
OUT
+ 0.008 40µV
23
23
PARD (Ripple and Noise)
RMS
Peak-to-Peak
0
0
_______mV
_______mV
0.003
0.030
450
µV
900
µV
25 Transient Recovery
(90% to 100% Full Load)
0 _______mV 0.075 9mV
Constant Current Tests
26 Current Programming
and Readback
Low Current (0.500A)
Front Panel Display
GPIB Readback
0.4896
I
OUT
- 0.012
I
OUT
- 0.012
________A
________A
________A
0.5104
I
OUT
+ 0.012
I
OUT
+ 0.012
221
µA
221
µA
221
µA
High Current (10.00A)
Front Panel Display
GPIB Readback
9.9815
I
OUT
- 0.031
I
OUT
- 0.031
________A
________A
________A
10.0185
I
OUT
+ 0.031
I
OUT
+ 0.031
4.1mA
4.1mA
4.1mA
27 Load Effect (3.30A)
I
OUT
- 0.0053 ________mA I
OUT
+ 0.0053 2µA
27 Source Effect (10.0A)
I
OUT
- 0.003 ________mA I
OUT
+ 0.003 4µA
27 PARD (Ripple and Noise)
0 ________mA 0.005 750
µA