Service manual
110
Performance Test Record - Agilent 6033A POWER SUPPLY (Page 2 of 2)
MODEL Agilent 6033A
Report No.______________ Date_____________________
Page
No.
Test Description
Minimum Spec. Results
*
Maximum
Spec.
Measurement
Uncertainty
Constant Voltage Tests
21 Voltage Programming
and Readback
Low Voltage (0.100V)
Front Panel Display
GPIB Readback
0.091
V
OUT
- 0.011
V
OUT
- 0.006
________V
________V
________V
0.109
V
OUT
+ 0.011
V
OUT
+ 0.006
3
µV
3
µV
3
µV
High Voltage (20.00V)
Front Panel Display
GPIB Readback
19.984
V
OUT
- 0.025
V
OUT
- 0.020
________V
________V
________V
20.016
V
OUT
+ 0.025
V
OUT
+ 0.020
300
µV
300
µV
300
µV
22 Load Effect (7.0V)
V
OUT
- 0.0027 _______mV V
OUT
+ 0.0027 20µV
22 Source Effect (20.00V)
V
OUT
- 0.003 _______mV V
OUT
+ 0.003 20µV
23
23
PARD (Ripple and Noise)
RMS
Peak-to-Peak
0
0
_______mV
_______mV
0.003
0.030
450
µV
900
µV
25 Transient Recovery
(90% to 100% Full Load)
0 _______mV 0.050 6mV
Constant Current Tests
26 Current Programming
and Readback
Low Current (1.000A)
Front Panel Display
GPIB Readback
0.9785
I
OUT
- 0.023
I
OUT
- 0.028
________A
________A
________A
1.0210
I
OUT
+ 0.023
I
OUT
+ 0.028
400
µA
400
µA
400
µA
High Current (30.00A)
Front Panel Display
GPIB Readback
29.935
I
OUT
- 0.155
I
OUT
- 0.115
________A
________A
________A
30.065
I
OUT
+ 0.155
I
OUT
+ 0.115
12mA
12mA
12mA
27 Load Effect (10.0A)
I
OUT
- 0.010 ________mA I
OUT
+ 0.010 19µA
27 Source Effect (30.0A)
I
OUT
- 0.009 ________mA I
OUT
+ 0.009 19µA
27 PARD (Ripple and Noise)
0 ________mA 0.015 2
µA