Operator`s manual

20 100 Operator’s Manual
3 Setting Up the 100
Setting up the Water Bath
1 Ensure the 100 is not plugged in.
2 Place the water bath on the 100 with the lower quick-connect fitting
positioned on the left side of the apparatus. Use the locator holes on
the water bath and the locator pins on the 100 to ensure proper
placement.
3 Locate the two tubes with quick-connect fittings and insert them into
the appropriate quick-connect fittings on the 100 water bath (see
Figure 1, “Water Bath Connections,” below).
Figure 1 Water Bath Connections
4 Ensure the cover is in place on the water bath with the small hole
toward the front. Ensure the disintegration beakers are not in place in
the cover.
5 Fill the water bath through any one of the three beaker holes in the
cover.
6 Check the water bath and all connecting tubing for leaks. When
satisfied that no leaks are present, proceed to “Installing the
Temperature Probe” on page 22.
NOTE
Initially fill the water bath no higher than 2.5 inches (6.4 cm) from the
top, to allow for displacement by the beakers when they are inserted.