Technical data

The 6060B and 6063B each provides
one load input. This is more conve-
nient for single input applications
than a mainframe product.
These electronic loads are particu-
larly suited for the lab bench.
Entering commands manually using
the front panel keypad is simpler
because the channel does not need
to be specified, as in a mainframe
configuration. The keypad entry is
further simplified because these
products do not have the download-
able LIST feature of the N3300A
Series, which helps to maximize
production throughput. Extensive
protection is included to help
protect your valuable prototypes
under test. This includes overvolt-
age, overcurrent, overtemperature,
overpower, and reverse polarity.
These loads are suitable for
manufacturing test systems where
maximizing speed is not critical.
They use industry standard
SCPI instructions, and also have
VXIPlug&Play drivers to simplify
system design. For the greatest
speed and accuracy in programming
and measurement, see the N3300A
Series of dc electronic loads.
Agilent dc Electronic Loads maximize throughput with real-life loading conditions
Specifications
6060B 6063B
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
82
Cost-effective for single input applications
Convenient optional front panel input connection
6060B and 6063B
Single-Input: 250 W to 300 W
Amperes 0 to 60 A 0 to 10 A
Volts 3 to 60 V 3 to 240 V
Maximum power (at 40° C) 300 W 250 W
Constant current mode
Ranges 0 to 6 A, 0 to 60 A 0 to 1 A, 0 to 10 A
Accuracy 0.1% ±75 mA 0.15% ±10 mA
Regulation 10 mA 8 mA
Constant voltage mode
Accuracy 0.1% ±50 mV 0.12% ±120 mV
Regulation (w/remote sense) 10 mV 10 mV
Constant resistance mode 0.033 to 1.0 0.20 to 24.0
Ranges 1 to 1,000 24 to 10,000
10 to 10,000 240 to 50,000
Accuracy
1 : 0.8% ±8 m(with
6 A at input) 24 : 0.8% ±200 m(with
1 A at input)
1 K: 0.3% ±8 mS (with 6 V at input) 10 K/: 0.3% ±0.3 mS (with 24 V at input)
10 K
: 0.3% ±8 mS (with
6 V at input) 50 K
: 0.3% ±0.3 mS (with
24 V at input)
Transient generator
Frequency range 0.25 Hz to 10 kHz 0.25 Hz to 10 kHz
Accuracy 3% 3%
Duty cycle range 3 to 97% (0.25 Hz to 1 kHz) 3 to 97% (0.25 Hz to 1 kHz)
6 to 94% (1 to 10 kHz) 6 to 94% (1 to 10 kHz)
Accuracy 6% of setting ±2% 6% of setting ±2%
Current level high range 60-A range: 10-A range:
Accuracy 0.1% ±350 mA 0.18% ±50 mA
Current level low range 6-A range: 1-A range:
Accuracy 0.1% ±80 mA 0.18% ±13 mA
Voltage level 3 to 60 V 3 to 240 V
Voltage level accuracy 0.1% ±300 mV 0.15% ±1.1 V
Readback specifications
Current readback accuracy 0.05% ±65 mA 0.12% ±10 mA
Voltage readback accuracy ±(0.05% + 45 mV) ±(0.1% + 150 mV)
Ripple and noise
(20-Hz to 10-MHz noise) 4 mA rms 1 mA rms
Current 40 mA peak-to-peak 10 mA peak-to-peak
Voltage 6 mV rms 6 mV rms