Technical data

80
Multiple-Input: 150 W to 600 W (Continued)
Table A-2 (Continued)
Supplemental Characteristics
Table A-2 lists the supplemental
characteristics, which are not warranted
but are descriptions of typical perfor-
mance determined either by design
or type testing.
Notes
2
Applies to all ranges.
N3302A N3303A N3304A N3305A N3306A N3307A
Programmable short
66 mmax. 200 mmax. 33 mmax. 33 mmax. 17 mmax. 33 mmax.
40 m 100 m 20 m 25 m 12 m 20 m
typical typical typical typical typical typical
Programmable open 20 kΩ≥80 kΩ≥20 kΩ≥20 kΩ≥20 kΩ≥80 k
Command processing time
Using discrete commands 3 ms 3 ms 3 ms 3 ms 3 ms 3 ms
Using List commands 1 ms 1 ms 1 ms 1 ms 1 ms 1 ms
List dwell characteristics
Range 0 - 10 s 0 - 10 s 0 - 10 s 0 - 10 s 0 - 10 s 0 - 10 s
Resolution 1 ms 1 ms 1 ms 1 ms 1 ms 1 ms
Accuracy 5 ms 5 ms 5 ms 5 ms 5 ms 5 ms
Measurement time
1000 samples (default) 20 ms 20 ms 20 ms 20 ms 20 ms 20 ms
(with specified (with specified (with specified (with specified (with specified (with specified
measurement measurement measurement measurement measurement measurement
accuracy) accuracy) accuracy) accuracy) accuracy) accuracy)
200 samples 10 ms 10 ms 10 ms 10 ms 10 ms 10 ms
(with <6% (with <6% (with <6% (with <6% (with <6% (with <6%
additional additional additional additional additional additional
fixed offset) fixed offset) fixed offset) fixed offset) fixed offset) fixed offset)
100 samples 9 ms 9 ms 9 ms 9 ms 9 ms 9 ms
(with <10% (with <10% (with <10% (with <10% (with <10% (with <10%
additional additional additional additional additional additional
fixed offset) fixed offset) fixed offset) fixed offset) fixed offset) fixed offset)
20 points 7 ms 7 ms 7 ms 7 ms 7 ms 7 ms
(with <30% (with <30% (with <30% (with <30% (with <30% (with <30%
additional additional additional additional additional additional
fixed offset) fixed offset) fixed offset) fixed offset) fixed offset) fixed offset)
<20 points 7 ms 7 ms 7 ms 7 ms 7 ms 7 ms
(with >30% (with >30% (with >30% (with >30% (with >30% (with >30%
additional additional additional additional additional additional
fixed offset) fixed offset) fixed offset) fixed offset) fixed offset) fixed offset)
Ripple and noise (20 Hz - 10 MHz)
Current (rms/peak to peak)
2 mA/20 mA 1 mA/10 mA 4 mA/40 mA 4 mA/40 mA 6 mA/60 mA 2 mA/20 mA
Voltage (rms) 5 mV
rms
12 mV
rms
6 mV
rms
10 mV
rms
8 mV
rms
10 mV
rms
External analog programming
Voltage Programming Accuracy
2
0.5% + 12 mV 0.5% + 48 mV 0.5% + 12 mV 0.5% + 30 mV 0.5% + 12 mV 0.5% + 30 mV
Current Programming Accuracy
2
0.25% + 4.5 mA 0.25% + 1.5 mA 0.25% + 9 mA 0.25% + 9 mA 0.25% + 18 mA 0.25% + 4.5 mA
External monitor ports
Voltage Monitor Accuracy
0.25% + 12 mV 0.25% + 48 mV 0.25% + 12 mV 0.25% + 30 mV 0.25% + 12 mV 0.25% + 30 mV
Current Monitor Accuracy
0.1% + 4.5 mA 0.1% + 1.5 mA 0.1% + 9 mA 0.1% + 9 mA 0.1% + 18 mA 0.1% + 4.5 mA
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
Agilent dc Electronic Loads maximize throughput with real-life loading conditions
3