Technical data
51
Supplemental Characteristics
for all model numbers
dc Floating Voltage: Output terminal can
be floated up to ±60 Vdc from chassis
ground
Remote Sensing: Up to half the rated
output voltage can be dropped in each
load lead. The drop in the load leads
subtracts from the voltage available at
the load.
Command Processing Time: Average
time required for the output voltage
to begin to change following receipt of
digital data is 20 ms for power supplies
connected directly to the GPIB.
Modulation: (analog programming of
output voltage and current):
Input Signal: 0 to –5 V for voltage,
and 0 to +5 V for current.
Input Impedance: 30 kΩ or greater.
ac Input (47 to 63 Hz): 180 to 235 Vac
(line-to-line 3 phase) 36 A rms maximum
worst case, 28 A rms nominal;
360 to 440 Vac, 18 A rms maximum
worst case, 14 A rms nominal.
(Maximum line current includes 5%
unbalanced phase voltage condition).
Input Power: 9000 VA and 7950 W
maximum; 175 W at no load.
Size: 425.5 mm W x 221.5 mm H x
674.7 mm D (16.75 in x
8.75 in x 25.56 in).
See page 104 for more details.
Warranty Period: Three years.
Single-Output: 6600 W GPIB (Continued)
Power Products Catalog 2002-2003
For more detailed specifications see the product manual at
www.agilent.com/find/power
Agilent Performance dc Power Supplies speed and accuracy for test optimization
Specifications
(at 0
˚
to 55
˚
C unless
otherwise specified)
6690A 6691A 6692A
Supplemental Characteristics (Non-warranted characteristics determined by design that are
useful in applying this product)
Ripple and noise constant current mode
from 20 Hz to 20 MHz
rms 200 mA 50 mA 30 mA
Average programming resolution
Voltage 4.1 mV 8.1 mV 16 mV
Current 118.5 mA 59 mA 30 mA
OVP 90 mV 170 mV 330 mV
Output voltage programming
response time (excludes 45 ms 60 ms 100 ms
command-processing time)
Full-load progrmming rise or fall time
(10 to 90% or 90 to 10%, resistive load)
Output common-mode
noise current rms 3 mA 3.5 mA 4 mA
(to signal-ground peak-to-peak 20 mA 20 mA 25 mA
binding post)
Ordering Information
Opt 208 180 to 235 Vac, 3 phase,
47 to 63 Hz
Opt 400 360 to 440 Vac, 3 phase,
47 to 63 Hz
Opt 602 Two Bus Bar Spacers for
paralleling power supplies
(p/n 5060-3514)
* Opt 908 Rack-mount Kit (p/n 5062-3977
and p/n 5063-9212)
* Opt 909 Rack-mount Kit with Handles
(p/n 5063-9221 and p/n 5063-9219).
Opt 0L2 Extra Standard
Documentation Package
Opt 0B3 Service Manual
Opt 0B0 No Documentation Package
* Support rails required
Accessories
p/n 5065-6935 Replacement fuse kit for
360-440 Vac line.
p/n 5065-6934 Replacement fuse kit for
180-235 Vac line.
E3663AC Support rails for Agilent
rack cabinets.
p/n 5080-2148 Serial link cable 2m (6.6 ft.)