D Single-Mode/Multimode Module Performance TestsOptional Test VI: E6007A Visual Fault Finder
266 E6000C Mini-OTDR Users Guide, E0302
Optional Test VI: E6007A Visual
Fault Finder Submodule
General
Make sure that all equipment has warmed up, and all
connectors are clean.
Make sure that all patchcords are fixed to the table, and
will not move during measurements.
Optional Test of Output Power Level (CW)
Connect equipment 1 Connect the equipment as shown in Figure 89.
Table 6 Visual Fault Finder: Test Equipment Required
Instrument or Accessories qty
Recommended
Agilent Model
Required Characteris-
tic
Alternative Models
Lightwave Multimeter
Meter
Optical Power Sensor
Patchcord (HMS10/
HMS10, 9/50
µm, SM)
connector interface
Optical Spectrum Analyzer
1
1
1
2
1
8163A
81101AC
81000AI
625-645nm
8164A, 8166A