Specifications
Installation 2
Hardware Installation 43
Verifying High Mass System Performance
Refer to the 5975 Series MSD Hardware Manual.
To check high mass performance, the system makes 1 injection of the high
mass checkout sample and analyzes the results.
For high mass systems, verify high mass system performance as follows:
1 Load the high mass checkout method, PFHT.M, and resolve any
configuration conflicts.
2 Load the high mass checkout sample into vial 2.
3 When the GC and MSD are ready, select Checkout/High Mass Check from
the Intrument view.
4 Verify that the rms signal-to-noise ratio meets the specifications
described in the 5975 Series MSD Hardware Manual.