Specifications

28 Hardware Installation
2 Installation
Configuring the GC
1 If there is an ALS, install it now.
Follow the instructions provided with the ALS.
2 Install the Merlin Microseal (optional). Follow the instructions provided
with the Merlin Microseal.
3 Connect the GC power cord to the appropriate electrical outlet.
4 Turn on the GC and verify that the GC passes self-test.
Keep the GC oven turned off.
5 Configure the column parameters for the MSD column that is installed in
the GC.
On the GC keypad, select [Config] [Col 1] and configure the column for the
GC Series that you have for the following parameters:
6 Increase the supply pressure of the carrier gas to 50 to 60 psi. (This may
be set higher, depending on the GC you are using).
7 Set the column flow rate to 5 mL/min and split flow to 200 mL/min.
Do not use syringes with tapered needles with a Microseal.
Length (m)
Diameter (ยต)
Film thickness
Inlet
Detector
Vacuum correct
Pres correct