Installation manual

Operating in Electron Impact (EI) Mode 3
5975 Series MSD Operation Manual 69
To set the GC/MSD interface temperature from the ChemStation
You can also use the Local Control Panel to perform this task. See “Operating
the MSD from the LCP” .
Procedure
1 Select View>Instrument Control.
2 Select Instrument>GC Edit Parameters.
3 Click the Aux icon to edit the interface temperature (Figure 16).
4 Check the heater On and type the setpoint in the Value °C column.
The typical setpoint is 280 °C. The limits are 0 °C and 350 °C. A setpoint
below ambient temperature turns off the interface heater.
Figure 16 Setting the interface temperature