3 Operating in Electron Impact (EI) Mode
4 To close the screen, click:
• Apply to send the new temperature setpoints to the MSD.
• OK to change the currently loaded tune file but not download anything to
the MSD (use Apply).
• Cancel to exit the panel without changing the currently loaded tune file
or downloading anything to the MSD.
5 When the Save MS Tune File dialog box appears, either click OK to save your
changes to the same file or type a new file name and click OK.
Table 12 Recommended temperature settings
EI operation PCI operation NCI operation
MS Source 230 250 150
MS Quad 150 150 150