3 Operating in Electron Impact (EI) Mode
4 To close the screen, click:
Apply to send the new temperature setpoints to the MSD.
OK to change the currently loaded tune file but not download anything to
the MSD (use Apply).
Cancel to exit the panel without changing the currently loaded tune file
or downloading anything to the MSD.
5 When the Save MS Tune File dialog box appears, either click OK to save your
changes to the same file or type a new file name and click OK.
Table 12 Recommended temperature settings
EI operation PCI operation NCI operation
MS Source 230 250 150
MS Quad 150 150 150