66 5975 Series MSD Operation Manual
3 Operating in Electron Impact (EI) Mode
To set monitors for MSD temperature and vacuum status
A monitor displays the current value of a single instrument parameter. They
can be added to the standard instrument control window. Monitors can be set
to change color if the actual parameter varies beyond a user-determined limit
from its setpoint.
Procedure
1 Select MS Monitors from the Instrument menu.
2 In the Edit MS Monitors box, under Type, select Zone.
3 Under Parameter, select MS Source and click Add.
4 Under Parameter, select MS Quad and click Add.
5 Under Parameter, select Foreline (or TurboSpd) and click Add.
6 Select any other monitors you want and Add them.
7 Click OK. The new monitors will be stacked on top of each other in the lower
right corner of the Instrument Control window. They must be moved for you
to see them all.
8 Click and drag each monitor to the desired position. See Figure 14 for one
way of arranging the monitors.
9 To make the new settings part of the method, select Save from the Method
menu.
Figure 14 Arranging monitors