Installation manual
62 5975 Series MSD Operation Manual
3 Operating in Electron Impact (EI) Mode
Controlling Column Flow
Carrier gas flow is controlled by head pressure in the GC. For a given head 
pressure, column flow will decrease as the GC oven temperature increases. 
With electronic pneumatic control (EPC) and the column mode set to Constant 
Flow, the same column flow is maintained regardless of temperature.
The MSD can be used to measure actual column flow. You inject a small 
amount of air or other unretained chemical and time how long it takes to 
reach the MSD. With this time measurement, you can calculate the column 
flow. See page 73.