Operating in Electron Impact (EI) Mode 3
5975 Series MSD Operation Manual 53
LCP Status Messages
The following messages may be displayed on the LCP to inform you of the
status of the MSD system. If the LCP is currently in Menu mode, cycle through
the menus to return to Status mode.
ChemStation Loading <timestamp>
The Agilent MSD Productivity ChemStation software is starting up.
Executing <type>tune
A tuning procedure is in progress (type = QuickTune or Autotune).
Instrument Available <timestamp>
The Agilent MSD Productivity ChemStation software is not running.
Loading Method <method name>
Method parameters are being sent to the MSD.
Loading MSD Firmware
The MSD’s firmware is being initialized.
The following messages alternately appear on the LCP if the MSD does NOT
complete its bootup sequence properly:
Server not Found
Check LAN Connection
Seeking Server
Bootp Query xxx
These messages indicate that the MSD has not received its unique IP address
from the Agilent Bootp Service. If the messages persist after you have logged
onto your account on the GC/MSD ChemStation, consult the Troubleshooting
section of the Software Installation manual.
NOTE
No messages will be displayed if an online instrument session is not currently running on
the GC/MSD ChemStation.