Installation manual
Introduction 1
5975 Series MSD Operation Manual 21
Mass spectrometer automated shutoff 
valves closed
Some mass spectrometers are equipped with 
automated diffusion pump shutoff valves. In these 
instruments, deliberate operator action or various 
failures can cause the shutoff valves to close. 
Shutoff valve closure does not shut off the flow of 
carrier gas. As a result, hydrogen may slowly 
accumulate in the mass spectrometer.
Mass spectrometer manual shutoff 
valves closed
Some mass spectrometers are equipped with 
manual diffusion pump shutoff valves. In these 
instruments, the operator can close the shutoff 
valves. Closing the shutoff valves does not shut off 
the flow of carrier gas. As a result, hydrogen may 
slowly accumulate in the mass spectrometer.
GC off  A GC can be shut down deliberately. It can also be 
shut down accidentally by an internal or external 
failure. Different GCs react in different ways. If a 
6890 GC equipped with Electronic Pressure Control 
(EPC) is shut off, the EPC stops the flow of carrier 
gas. If the carrier flow is not under EPC control, the 
flow increases to its maximum. This flow may be 
more than some mass spectrometers can pump 
away, resulting in the accumulation of hydrogen in 
the mass spectrometer. If the mass spectrometer is 
shut off at the same time, the accumulation can be 
fairly rapid.
Power failure  If the power fails, both the GC and mass 
spectrometer shut down. The carrier gas, however, 
is not necessarily shut down. As described 
previously, in some GCs a power failure may cause 
the carrier gas flow to be set to maximum. As a 
result, hydrogen may accumulate in the mass 
spectrometer.
Table 4 Hydrogen accumulation mechanisms (continued)
Mechanism Results










