Installation manual

136 5975 Series MSD Operation Manual
5 General Maintenance
To reinstall the EI ion source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Pliers, long-nose (8710-1094)
Procedure
1 Slide the ion source into the source radiator (Figure 33).
2 Install and hand tighten the source thumbscrews. Do not overtighten the
thumbscrews.
3 Connect the ion source wires as shown in “To close the analyzer chamber” .
Close the analyzer chamber.