5975 Series Site Prep.book Page 2 Thursday, January 31, 2008 9:47 AM Notices © Agilent Technologies, Inc. 2008 Warranty No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws.
5975 Series Site Prep.
5975 Series Site Prep.
975 Series Site Prep.book Page 5 Thursday, January 31, 2008 9:47 AM 5975 Series Mass Selective Detectors Site Preparation Guide 1 General Information Instrument Identification 6 Important Safety Warnings 7 Hydrogen Safety 9 Chemical Safety 10 Safety and Regulatory Certifications 11 Cleaning 14 Recycling the Product 14 This section provides information on how to identify specifically what type of MSD you are working with as well as how to properly clean and dispose of the instrument.
5975 Series Site Prep.book Page 6 Thursday, January 31, 2008 9:47 AM General Information Instrument Identification Each 5975 Series Mass Selective Detector (MSD) is identified by a unique 10-character serial number. This serial number is located on a label on the lower left side near the front of the instrument. On CI upgrades, a second serial number is located on the flow module.
5975 Series Site Prep.book Page 7 Thursday, January 31, 2008 9:47 AM General Information Important Safety Warnings Before moving on, there are several important safety notices that you should always keep in mind when using the 5975 Series Mass Selective Detector.
5975 Series Site Prep.book Page 8 Thursday, January 31, 2008 9:47 AM General Information The use of incorrect or makeshift fuses or the short-circuiting of fuse holders creates a shock hazard for the operator and can damage the instrument. Replace fuses only with fuses of identical current rating and type. Excessive fluctuations in the line voltage can create a shock hazard and can damage the instrument. Make sure the supply voltage does not fluctuate more than +5% or -10% from the rated voltage.
5975 Series Site Prep.book Page 9 Thursday, January 31, 2008 9:47 AM General Information Hydrogen Safety Hydrogen gas may be used as carrier gas, and/or as fuel for certain GC detectors. When mixed with air, hydrogen can form explosive mixtures. A detailed discussion of hydrogen safety appears in Chapter 1 of the Agilent 5975 MSD Troubleshooting and Maintenance Manual. Read this material before using hydrogen with the MSD.
5975 Series Site Prep.book Page 10 Thursday, January 31, 2008 9:47 AM General Information Chemical Safety WA RN ING 10 The foreline pump exhaust and split vent exhaust will contain traces of the chemicals you are analyzing. These could potentially be toxic. Vent the foreline pump exhaust and split vent exhaust outside your laboratory or into a fume hood. Be sure to comply with all local environmental regulations.
5975 Series Site Prep.book Page 11 Thursday, January 31, 2008 9:47 AM General Information Safety and Regulatory Certifications The 5975 Series MSD conforms to the following safety standards: • Canadian Standards Association (CSA): CAN/CSA–C22.2 No.
5975 Series Site Prep.book Page 12 Thursday, January 31, 2008 9:47 AM General Information Symbols Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Agilent Technologies assumes no liability for the customer’s failure to comply with these requirements. See accompanying instructions for more information.
5975 Series Site Prep.book Page 13 Thursday, January 31, 2008 9:47 AM General Information Electromagnetic compatibility This device complies with the requirements of CISPR 11. Operation is subject to the following two conditions: • This device may not cause harmful interference. • This device must accept any interference received, including interference that may cause undesired operation.
5975 Series Site Prep.book Page 14 Thursday, January 31, 2008 9:47 AM General Information Cleaning To clean the unit, disconnect the power and wipe down with a damp, lint-free cloth. Recycling the Product For recycling, contact your local Agilent sales office.
5975 Series Site Prep.book Page 15 Thursday, January 31, 2008 9:47 AM 5975 Series Mass Selective Detectors Site Preparation Guide 2 Site Preparation Overview 16 Space and Weight Requirements 20 Communications Requirements 22 Electrical Requirements 23 Air Conditioning and Environmental Requirements 28 Carrier and Reagent Gas Requirements 31 Laboratory Supply Requirements 33 Receiving the System 37 Installation and Verification 39 This section describes how to properly prepare your site for a new MSD.
5975 Series Site Prep.book Page 16 Thursday, January 31, 2008 9:47 AM Site Preparation Overview Before the 5975 Series MSD system can be installed, the site must be properly prepared. Site preparation includes, but is not limited to, ensuring that adequate facilities are available. Among the site requirements are: • Adequate space is available for the MSD system. • A suitable supporting bench is available. • Adequate electrical power is available at the correct voltages and frequencies.
5975 Series Site Prep.book Page 17 Thursday, January 31, 2008 9:47 AM Site Preparation • Complying with all local laws (codes, ordinances, and regulations) for mechanical, building, and electrical distribution systems, hazardous waste disposal, and chemical storage. Compliance must exist prior to installation.
5975 Series Site Prep.
5975 Series Site Prep.book Page 19 Thursday, January 31, 2008 9:47 AM Site Preparation Other documentation Additional information is contained in the following documentation: • 5975 Series MSD Troubleshooting and Maintenance Manual • 5975 Series MSD Operation manual • 6890 Series GC manuals • GC accessories (autosampler, etc.) manuals • G1701EA MSD Productivity ChemStation software manuals and online help • The appropriate sensitivity specifications for your instrument.
5975 Series Site Prep.book Page 20 Thursday, January 31, 2008 9:47 AM Site Preparation Space and Weight Requirements Table 2 lists dimension and weight information for the MSD and related components. Your site must have enough bench space for the MSD, GC, data system, and accessories (Figure 1). In addition, there must be sufficient space around the system for ventilation and maintenance access.
5975 Series Site Prep.
5975 Series Site Prep.book Page 22 Thursday, January 31, 2008 9:47 AM Site Preparation Communications Requirements Telephone A telephone with a cord long enough to be used at the computer will allow the MSD operator to communicate with Agilent Technologies support personnel. Site LAN network If you intend to connect your MSD system to your site’s LAN network, you must have an additional shielded twisted pair network cable.
5975 Series Site Prep.book Page 23 Thursday, January 31, 2008 9:47 AM Site Preparation Electrical Requirements You are responsible for providing appropriate electrical power and power outlets for all of the components in your 5975 Series MSD system.
5975 Series Site Prep.book Page 24 Thursday, January 31, 2008 9:47 AM Site Preparation Power configurations Electrical power for the MSD must be single-phase. The neutral wire cannot be used for safety grounding. The ground wire should carry zero current except for ground-fault current or static electric discharge. The entire system should share an isolated, noise-free electrical ground.
5975 Series Site Prep.book Page 25 Thursday, January 31, 2008 9:47 AM Site Preparation Power requirements Table 4 lists the power requirements for the 5975 Series MSD and related equipment. Extra power capacity for future additions is recommended. Each product listed requires a dedicated circuit. The gauge controller is powered by the MSD. The MSD, GC, and data system must each have a separate circuit breaker. All of the equipment must share a common ground.
5975 Series Site Prep.book Page 26 Thursday, January 31, 2008 9:47 AM Site Preparation Table 4 Power requirements (continued) Product Line voltage Maximum continuous power consumption Supply circuit rating Outlets required ChemStation computer 100-120 or 220-240 VAC+5/-10%, 50/60 Hz ± 5% 1000 VA 15 A 3 to 5 * The MSD operates on either voltage range.
5975 Series Site Prep.book Page 27 Thursday, January 31, 2008 9:47 AM Site Preparation WA RN ING Do not use extension cords with the MSD or GC. Extension cords normally are not rated to carry enough power and can be a safety hazard. NO TE Maintain easy access to the power cords so they can be disconnected for maintenance.
5975 Series Site Prep.book Page 28 Thursday, January 31, 2008 9:47 AM Site Preparation Air Conditioning and Environmental Requirements Air conditioning considerations include temperature, humidity, airborne dust, and exhaust venting. Each of these is considered in more detail in the following material.
5975 Series Site Prep.book Page 29 Thursday, January 31, 2008 9:47 AM Site Preparation Airborne dust Agilent Technologies recommends a maximum airborne particle density of 55 μg/m3. If you suspect your site exceeds this limit, contact your local Agilent Technologies Customer Service Organization. Customer Engineers with special training and equipment can test for airborne particle density. They can offer suggestions for reducing airborne dust.
5975 Series Site Prep.book Page 30 Thursday, January 31, 2008 9:47 AM Site Preparation See your GC manuals for information on venting the split vent. Fume (exhaust) hood An auxiliary work space and fume hood are needed for some maintenance procedures.
5975 Series Site Prep.book Page 31 Thursday, January 31, 2008 9:47 AM Site Preparation Carrier and Reagent Gas Requirements Table 6 lists the specifications for the carrier and reagent gases. Pressures and flows are as measured at the bulkhead fittings on the back of the instrument, not at the gas supply. We recommend that additional shut-off valves be installed near the instruments.
5975 Series Site Prep.book Page 32 Thursday, January 31, 2008 9:47 AM Site Preparation Regulators, tubing, and fittings You must supply appropriate regulators for your carrier and reagent gas. • Regulators must be dual-stage. • Regulators must have stainless-steel (not elastomeric) diaphragms. • They must supply gas at the pressure specified. • They must have one outlet with 1/4-inch Swagelok fittings.
5975 Series Site Prep.book Page 33 Thursday, January 31, 2008 9:47 AM Site Preparation Laboratory Supply Requirements Laboratory supply requirements vary with applications. Cleaning solvents Cleaning tasks for the MSD require the following HPLC-grade (or better) solvents: • Methylene chloride • Isopropyl alcohol • Methanol • Acetone • Water Proper storage, handling, and disposal of these chemicals is required for personal and environmental safety.
5975 Series Site Prep.book Page 34 Thursday, January 31, 2008 9:47 AM Site Preparation Spare parts and consumables The supplies and parts listed in Table 7 are used in the operation and maintenance of a 5975 Series MSD system. Keeping these parts on hand can reduce system downtime related to instrument maintenance and repair.
5975 Series Site Prep.book Page 35 Thursday, January 31, 2008 9:47 AM Site Preparation Table 7 Maintenance supplies and consumables (continued) Chemical Consumables PFTBA (EI calibration fluid) 05971-60571 PFDTD (CI calibration fluid) 8500-8130 EVAL A (evaluation sample) 05990-60045 OFN (Octafluoronapthalene) 5188-5348 OFN 100 fg/µL 5188-5347 Benzophenone 8500-5440 PFHT 5188-5357 Foreline pump oil, P3, 0.
5975 Series Site Prep.book Page 36 Thursday, January 31, 2008 9:47 AM Site Preparation Table 7 Maintenance supplies and consumables (continued) 0.5-mm id, 85% Vespel 15% graphite, for 0.32-mm id columns 5062-3506 0.8-mm id, 85% Vespel 15% graphite, for 0.53-mm id columns 5062-3538 GC inlet 36 0.27-mm id, 90% Vespel 10% graphite, for 0.10-mm id columns 5062-3518 0.37-mm id, 90% Vespel 10% graphite, for 0.20-mm id columns 5062-3516 0.40-mm id, 90% Vespel 10% graphite, for 0.
5975 Series Site Prep.book Page 37 Thursday, January 31, 2008 9:47 AM Site Preparation Receiving the System When your MSD system is delivered, it is your responsibility to provide for removal of the shipping containers from the truck and their storage until installation. Contact your Agilent Technologies service representative as soon as your shipment arrives to arrange an installation date. Delivery and unloading The shipping containers are large and heavy. The largest container in the shipment is 121.
5975 Series Site Prep.book Page 38 Thursday, January 31, 2008 9:47 AM Site Preparation Storage It is your responsibility to store the containers until installation. If your site does not have adequate storage space, the containers may be stored at your expense in a bonded warehouse. Allow space for data system and accessory containers too. The environment in the storage area should be between 5 °C and 50 °C (41 °F and 122 °F), 20% to 80% relative humidity, non-condensing and non-corrosive.
5975 Series Site Prep.book Page 39 Thursday, January 31, 2008 9:47 AM Site Preparation Installation and Verification Installation Once the installation has begun, it should progress in a timely manner to completion. Delays due to inadequate site preparation could cause loss of instrument use during the warranty period. In extreme cases, Agilent Technologies may ask to be reimbursed for the additional time required to complete the installation.
5975 Series Site Prep.book Page 40 Thursday, January 31, 2008 9:47 AM Site Preparation Sensitivity specifications See the appropriate sensitivity specifications for your instrument type. See the Agilent Technologies Web site at http://www.agilent.com/chem. NO TE 40 SIM performance verification is not performed at installation and must be purchased if desired.
5975 Series Site Prep.book Page 41 Thursday, January 31, 2008 9:47 AM 5975 Series Mass Selective Detectors Site Preparation Guide A Power Cords This appendix shows the power cords available for the 5975 Series MSD. See page 26 for information about ensuring that the correct power cord is supplied with your system.