Specifications

70 5975 Series MSD Troubleshooting and Maintenance Manual
3 CI Troubleshooting
No reagent gas peaks in NCI
Reagent gases do not ionize in NCI; look for background ions instead
Verify tune parameters
If no background ions are visible, go back to methane PCI
No PFDTD calibrant peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
)
Verify tune parameters
Go back to methane PCI
No sample peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
)
Go back to methane PCI
Poor quality reagent gas (purity less than 99.99%)
Large peak at m/z 238 in NCI OFN spectrum
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
)
Find and fix your small air leak