50 5975 Series MSD Troubleshooting and Maintenance Manual
2 General Troubleshooting
Electrical leads to the detector are not connected correctly
HED power supply output cable failed
Electrical leads to the ion source are not connected correctly
Filament shorted to the source body
Temperature control disabled
One of the heater fuses has failed
MSD electronics are not working correctly
Temperature control fault
This indicates that something has gone wrong with the temperature control of
either the ion source or mass filter (quad) heater. The cause can be further
isolated by selecting Status/MS Temp Ctlr Status in the Tune and Vacuum
Control view. One of the following should be displayed as the cause:
Source temperature sensor is open
Source temperature sensor is shorted
Mass filter (quad) temperature sensor is open
Mass filter (quad) temperature sensor is shorted
No heater voltage (heater fuse has probably failed)
Heater voltage is too low
Temperature zone has timed out (heater failed, bad heater wiring, or loose
temperature sensor)
Problem with the temperature control electronics
Source heater is open
Source heater is shorted
Mass filter heater is open
Mass filter heater is shorted