Specifications

46 5975 Series MSD Troubleshooting and Maintenance Manual
2 General Troubleshooting
Mass filter heater failed (burned out or shorted to ground)*
Mass filter temperature sensor failed*
Source power cable is not connected to the sideboard*
MSD electronics are not working correctly
* These will cause an error message.
GC/MSD interface will not heat up
Incorrect setpoint(s)
Setpoint entered in wrong heated zone
GC/MSD interface has not had enough time to reach temperature setpoint
GC is off
GC experienced a fault and needs to be reset*
GC/MSD interface heater/sensor cable is not connected*
GC/MSD heater failed (burned out)*
GC/MSD sensor failed*
GC electronics are not working correctly*
* These will cause a GC error message. GC error messages are described in