Foreline pressure is too low
If the pressures you observe are below 20 mTorr, check for the following:
Column (carrier gas) flow is too low
Column plugged or crushed by an overtightened nut
Empty or insufficient carrier gas supply*
Bent or pinched carrier gas tubing*
Foreline gauge is not working correctly
* These could create a fault condition in the GC that would prevent the GC
from operating.
Analyzer chamber pressure is too low
If the pressures you observe are below 1 × 10
-6
Torr, check for the following:
Column (carrier gas) flow is too low
Column plugged or crushed by overtightened nut
Empty or insufficient carrier gas supply*
Bent or pinched carrier gas tubing*
* These could create a fault condition in the GC that would prevent the GC
from operating.
Gauge controller displays 9.9+9 and then goes blank
This indicates the pressure in the analyzer chamber is above 8 × 10
-3
Torr.
Solvent peak from an on-column injection
MSD has not had enough time to pump down
Excessive foreline pressure
Vacuum gauge has failed
Line voltage too low
Turbo pump is not working correctly
Diffusion pump fluid level is low or fluid is contaminated