2 General Troubleshooting
Mass Spectral Symptoms
This section describes symptoms you might observe in mass spectra. Some of
these symptoms will appear in the mass spectra of samples. Others you will
observe only in a tune report. Some of these symptoms have causes that can be
corrected by the operator. Others, however, require service by an Agilent
Technologies service representative.
Two symptoms listed under Chromatographic symptoms: If sensitivity is poor
and If repeatability is poor, also apply to mass spectra.
No peaks
Ion source cables not connected
Bad connections to or from the detector
HED power supply output cable has failed
Other electronics failure
Isotopes are missing or isotope ratios are incorrect
Peaks are too wide or too narrow
Scan speed is too high (scan mode)
Dwell time is too short (SIM mode)
Electron multiplier voltage is too high
Repeller voltage is too high
High background
Dirty ion source
High background
Pressure in the analyzer chamber is too high
Air leak
Contamination