Specifications

General Troubleshooting 2
5975 Series MSD Troubleshooting and Maintenance Manual 37
Poor sensitivity
Incorrect tuning
Tune file that does not match the type of analysis
Repeller voltage is too low
Incorrect temperatures (oven, GC/MSD interface, ion source, or mass filter)
Incorrect sample concentration
Leaking GC inlet*
Dirty GC inlet
Incorrect split ratio
Purge off time in splitless mode is too short
Excessive pressure in the MSD
Dirty ion source
Air leak
Poor filament operation
Detector (HED electron multiplier) is not working correctly
Incorrect mass filter polarity
* This could cause a fault condition in the GC that would prevent the GC
from operating.
Poor repeatability
Dirty syringe needle
Dirty GC inlet
Leaking GC inlet*
Injection is too large
Loose column connections
Variations in pressure, column flow, and temperature
Dirty ion source
Loose connections in the analyzer
Ground loops
* This could cause a fault condition in the GC that would prevent the GC
from operating.