Specifications

36 5975 Series MSD Troubleshooting and Maintenance Manual
2 General Troubleshooting
Baseline wanders
Insufficient carrier gas supply pressure*
Malfunctioning flow or pressure regulator*
Intermittent leak in the GC inlet*
* These could cause a fault condition in the GC that would prevent the GC
from operating.
Retention times for all peaks drift – shorter
Column has been shortened
Initial oven temperature was increased
Column is getting old
Retention times for all peaks drift – longer
Column flow has been reduced
Initial oven temperature was decreased
Active sites in the sample path
Leaks in the GC inlet*
* This could cause a fault condition in the GC that would prevent the GC
from operating.