Specifications
General Maintenance 4
5975 Series MSD Troubleshooting and Maintenance Manual 153
Maintaining the GC/MSD Interface
The GC/MSD interface requires no periodic maintenance
Rarely, the heater cartridge in the GC/MSD interface fails. In those cases, it is
necessary to replace the heater and sensor. This section contains procedures
for removing the heater and sensor and installing new ones. See Figure 28.
Figure 27 EI GC/MSD interface
MSD GC oven
Ionization chamber
Heater/sensor assembly
Analyzer chamber
Heater sleeve
Insulation