To disassemble the EI ion source
Materials needed
Gloves, clean, lint-free
Large (8650-0030)
Small (8650-0029)
Hex ball driver, 1.5 mm (8710-1570)
Hex ball driver, 2.0 mm (8710-1804)
Wrench, open-end, 10 mm (8710-2353)
Procedure
1 Remove the ion source. See the Operation Manual.
2 Remove the filaments (Figure 20).
3 Separate the repeller assembly from the source body. The repeller
assembly includes the source heater assembly, repeller, and related parts.
4 Remove the repeller.
5 Unscrew the interface socket. A 10-mm open-end wrench fits the flats on
the interface socket.
6 Remove the setscrew for the lenses.
7 Push the lenses out of the source body.