Specifications

General Maintenance 4
5975 Series MSD Troubleshooting and Maintenance Manual 117
To Reinstall a Micro-Ion Vacuum Gauge
Material needed
KF16 O-ring 0905-1463
Procedure
1 Place the KF16 O-ring in the groove on the analyzer chamber flange.
Replace it if it is worn or damaged.
2 Hold the gauge flange against the chamber flange with the O-ring. Push the
clamp over both flanges.
3 Insert the long screw, add the thumbnut, and tighten.
4 Attach the cable to the back of the gauge.