Specifications

General Maintenance 4
5975 Series MSD Troubleshooting and Maintenance Manual 109
To Refill the EI Calibration Vial
Materials needed
PFTBA (05971-60571)
Procedure
1 Stop any tuning or data acquisition.
2 Turn off the analyzer. There are several ways to do this:
In the Tune and Vacuum Control view, select MS OFF from the Execute
menu.
In the Instrument Control view in the Edit Parameters dialog box, select
MS OFF from the Execute menu.
3 If your MSD is equipped with a vacuum gauge, turn off the gauge.
4 Remove the analyzer window cover.
5 Loosen the calibration vial collar (Figure 13). Do not remove the collar.
6 Pull the calibration vial out. You may feel some resistance due to residual
vacuum.
Figure 13 Removing the EI calibration vial
Calibration vial
Collar