To operate the time base controls
The following exercise guides you through the time base keys, knobs, and
status line.
Turn the Time/Div knob and notice the change it makes to the status
line.
The Time/Div knob changes the sweep speed from 2 ns to 50 s (HP 54645A)
and from 5 ns to 50 s (HP 54645D) in a 1-2-5 step sequence, and the value is
displayed in the status line.
You can access the horizontal Vernier softkey by pressing the
Main/Delayed
key.
The Vernier softkey allows you to change the sweep speed in smaller
increments with the Time/Div knob. These smaller increments are
calibrated, which result in accurate measurements, even with the
vernier turned on.
You can check the sample rate by pressing the
Main/Delayed
key
when the instrument is running, during a single run, or when the
instrument is stopped. The sample rate is also displayed when the
acquisition stops or ends.
Turn the Delay knob and notice that its value is displayed in the status line.
The Delay knob moves the main sweep horizontally, and it pauses at
0.00 s, mimicking a mechanical detent. At the top and bottom of the
graticule is a solid triangle (
) symbol and an open triangle (
)
symbol. The
symbol indicates the trigger point and it moves with
the Delay knob. The
symbol indicates the time reference point. If
the time reference softkey is set to left, the
is located one graticule
in from the left side of the display. If the time reference softkey is set
to center, the
is located at the center of the display. The delay
number tells you how far the reference point
is located from the
trigger point
.