Specifications

To use glitch triggering
1
Press
Advancd
. Then select
Glitch
.
2
Select
Glitch Menu
.
3
Select the analog channel (A1 or A2) or digital channel (D0-D15)
source for the trigger by pressing the
Source
softkey until the desired
source is highlighted.
4
Select positive or negative polarity for the glitch to capture.
5
Select the time qualifier (<, less than a specified width; >, greater
than a specified width; or ><, within minimum and maximum limits)
for the glitch, and rotate the Entry knob or press the softkey to
increase the duration time.
Triggering HP 54645A/D Oscilloscopes