Specifications

Toggle each softkey and notice which keys cause the status line to change.
Channels 1 and 2 have a vernier softkey that allows the Volts/Div
knob to change the vertical step size in smaller increments. These
smaller increments are calibrated, which result in accurate
measurements, even with the vernier turned on.
To turn the channel off, either press
1
(for HP 54645A) or
A1
(for HP 54645D) a second time or press the left-most softkey.
Invert Operating Hint
Inversion affects how a channel is displayed, but it does not affect triggering.
The trigger does not see the inversion, and thus is not affected. If the trigger is
set to a rising edge, it remains rising after inversion.
Bandwidth Limit
The displayed bandwidth limit of the selected channel is 20 MHz, and is
available for channels 1 and 2.
Front-Panel Overview
To adjust oscilloscope vertical scaling
2-22