Specifications
5
Press
Time
, then press the
Freq
and
Period
softkeys.
You should measure the following:
Frequency 1 MHz, test limits are 995.9 kHz to 1.0042 MHz.
Period 1
µ
s, test limits are 995.9 ns to 1.0041
µ
s.
If the measurements are not within the test limits, see "Troubleshooting the
Oscilloscope."
6
Change the time mark generator to 10 ns, and change the time base to
5 ns/div. Adjust the trigger level to obtain a stable display.
7
Press
Time
, then press the
Freq
and
Period
softkeys.
You should measure the following:
Frequency 100 MHz, test limits are 98.61 MHz to 101.43 MHz.
Period 10 ns, test limits are 9.86 ns to 10.14 ns.
If the measurements are not within the test limits, see "Troubleshooting the
Oscilloscope."
Testing, Adjusting, and Troubleshooting
To verify horizontal
∆
t and 1/
∆
t accuracy
7-23