Specifications
4
Press
Single
.
Pressing
Single
arms the trigger circuit. When the trigger conditions
are met, data appears on the display, representing the data points that
the oscilloscope obtained with one acquisition. Pressing
the
Single
again re-arms the trigger circuit and erases the display.
Operating Hints
When Trigger Mode is set to Auto or Auto Lvl (auto-single mode), each
time is pressed the oscilloscope will wait for a valid trigger. If no
trigger is found after about 40 ms, the oscilloscope will force a trigger, acquiring
whatever signal is present. This is a handy mode to look for totally unknown
situations.
When Trigger mode set to Normal, pressing will arm the trigger
circuit and do a single acquisition when a valid trigger is seen.
You can use trace/pixel memories to compare several single-shot events. See
the "Saving and Recalling Configuration" section later in this chapter.
Single
Single
Run/Stop/Single/Autostore/Erase Operation
To capture a single event
4-35