User`s guide
53220A/53230A User’s Guide 285
53220A/53230A Error Messages A
+820
Model and Serial Numbers not
restored
Following replacement of the P500 processor
(controller) board or main measurement board,
the model number and/or serial number do not
match or have not been re-stored. Re-store the
model number and serial number as prompted
from the front panel. Refer to the service guide for
additional information.
+821
Controller and measurement
board model numbers do not
match
+822
Controller and measurement
board serial numbers do not
match
Calibration Errors
+701
Calibration error;
security defeated
Shorting the calibration secure jumper (CAL
ENABLE) when turning ON the instrument will
cause this error to occur indicating the security
password has been overwritten. See “Resetting
the Security Code to a Null” in the service guide
for details.
+702
Calibration error;
calibration memory is secured
A calibration cannot be performed when calibra-
tion memory is secured. See “To Unlock The
Counter For Calibration” and “To Lock the
Counter” in Chapter 2 of the service guide for
more information. Use the CAL:SEC:STAT ON
command to enter the security code using the
remote interface.
+703
Calibration error;
secure code provided was
invalid
The security code specified was invalid.
+704
Calibration error;
secure code too long
The security code specified is greater than 12
characters.
+705
Calibration error;
calibration aborted
An in-progress calibration was aborted (stopped)
by the user.
+706
Calibration error;
provided value out of range
A calibration value was entered that was outside
the expected range of valid inputs.
+707
Calibration error;
computed correction factor
out of range
If this error occurs, cycle power and run the instru-
ment self-test. Verify the calibration setup (instru-
ment settings, connections) and recalibrate the
counter. Contact Agilent if there is a self-test fail-
ure or the unit fails to calibrate.