53220A/53230A Error Messages A
-313
Calibration memory lost;
due to firmware revision
change
The instrument’s calibration data has been erased
following a instrument firmware update. Recali-
bration of the instrument is required.
-314
Save/recall memory lost;
memory corruption detected
The instrument state saved by the *SAV? com-
mand in non-volatile (flash) memory has been
lost.
-314
Save/recall memory lost;
due to firmware revision
change
The instrument state saved by the *SAV? com-
mand in non-volatile (flash) memory has been lost
due to a firmware update.
-315
Configuration memory lost;
memory corruption detected
User-preference settings such as I/O settings, the
reference oscillator setting, measurement time
out setting,... have been lost.
-315
Configuration memory lost;
due to firmware revision
change
User-preference settings such as I/O settings, the
reference oscillator setting, measurement time
out setting,... have been lost due to a firmware
update.
-330 Self-test failed
See the appended message for more information.
-350 Error queue overflow
An error has occurred; however, the instrument
error queue is full and the error was not recorded.
Query Errors
-410 Query IINTERRUPTED
The counter received a command before it was
finished responding to a previous query.
-420 Query UNTERMINATED
The computer attempts to read a query response
from the counter after having issued an incom-
plete query command.
-430 Query DEADLOCKED
The counter’s input and output buffers are full and
the instrument cannot continue.
-440
Query UNTERMINATED after
indefinite response
A query command was received in the same
string after a query (e.g. *IDN?) requesting an
indefinite length response.
Instrument Errors
+100 Network Error
A LAN network error has occurred or there is a
possible hardware failure. Run the instrument
self-test to determine if a hardware failure exists.
If so, contact Agilent.
+110 LXI mDNS Error