1 Preparation for Use
The following table summarizes battery operation:
Battery Care
When the battery is enabled and the instrument is not connected to AC
power, the battery discharges at 30% of full capacity per day. When the battery
is disabled and AC power is not connected, the battery discharges at 10% of
full capacity per month.
When storing the instrument without AC power applied, do not allow the
battery to discharge below 10%. The following equation can be used to
determine instrument storage time without AC power - and still enable the
instrument to be fully recharged:
months battery disabled * 10% + days battery enabled * 30%
= 90%
A fully discharged battery may need to be replaced if it has not been recharged
for six to 18 months.
Operating Time (typ): 3 hours (below +35 °C)
Standby Time (typ): 24 hours (below +35
°C, OCXO powered)
Recharge Time (typ): 4 hours to 100% capacity or 2 hours to 90% capacity
Temperature Range: 0
°C to +55 °C (operating) - battery charges below +35 °C
-10
°C to +60 °C (storage)
NOTE
If battery operation is in use above the maximum specified operating temperature, the
battery will shut down the instrument to preserve its use. AC power must be applied to
recover from a shut down caused by this condition.
NOTE
For a complete list of battery and all 53220A/53230A product specifications, refer to the
data sheet included on the 53210A/53220A/53230A Product Reference CD
(p/n 53220-13601), or on the Web at:
www.agilent.com/find/53220A
www.agilent.com/find/53230A