User`s guide

53220A/53230A User’s Guide 217
53220A/53230A Math, Graphing, and Data Logging 6
|MINimum|MAXimum|DEFault}
CALCulate2:TRANsform:HISTogram:RANGe:UPPer? [{MINimum
|MAXimum|DEFault}] (query form)
value - directly specifies the histogram lower and upper range values. The
value ranges are:
-1.0E+15 to -1.0E-15, 0.0, 1.0E-15 to 1.0E+15
The default lower and upper ranges and the ranges set following a reset
(*RST) or instrument preset (SYSTem:PRESet) are 0.0.
When setting the ranges, both a lower range and an upper range must be spec-
ified. Any of the following steps will prevent a “settings conflict” error
(e.g. lower range greater than upper range) due to the order in which ranges
are set:
setting the upper range first
setting both ranges on the same line programmatically
enabling the histogram after the ranges are set
setting the ranges automatically
Setting the Bin Ranges to Min and Max
With AutoRange Off, the Set to Min/Max softkey is another method for setting the
histogram bin ranges (lower and upper). Depending on the instrument state,
Set to Min/Max will set the ranges as follows: