User`s guide

53220A/53230A User’s Guide 193
53220A/53230A Math, Graphing, and Data Logging 6
Figure 6-1. Enabling the Math Operations.
Enabling the CALCulate1 Subsystem
Individual math operations require the CALCculate1 subsystem and the spe-
cific math operation to be enabled before the operation is performed. The
command used to enable the CALCulate1 subsystem is:
CALCulate[1][:STATe] {OFF|ON}
CALCulate[1][:STATe]? (query form)
ON enables the subsystem, OFF disables the subsystem. With the subsystem
disabled, measurement data is sent directly to the output buffer and/or read-
ing memory with no math operations performed, regardless of any math
operations enabled.
reading memory
CALCulate1:AVERage
(statistics)
CALCulate1:SCALe
(scaling and offset)
CALCulate1:LIMit
(limit checking)
CALCulate2:TRANsform:HISTogram
(histograms)
m
easurement
data
CALCulate1:STATe ON
CALCulate1:SCALe: STATe ON
CALCulate2:TRANsform:HISTogram:STATe ON
CALCulate1:LIMit:STATe ON
CALCulate1:AVERage:STATe ON
INITiate:IMMediate
READ?
CALCulate1:SMOothing
(moving average)
CALCulate1:SMOothing:STATe ON
outpu
t
buffe
r
readings initiated by INITiate:IMMediate are sent to reading memory.
readings initiated by READ? are sent to reading memory and to the output buffer