User`s guide
172 53220A/53230A User’s Guide
5 Triggering and Gating
CONF:TINT (@1),(@2) // time int between ch1/ch2
INP1:LEV1 2 // set start level (event)
INP1:SLOP1 POS // set start event polarity
INP2:LEV1 2 // set stop level (event)
INP2:SLOP1 NEG // set stop event polarity
SENS:TINT:GATE:POL POS // set gate signal polarity
SENS:TINT:GATE:SOUR EXT // external gate source
Enabling Gate Signals on the ‘Gate In/Out’ BNC
For timing and synchronization with other instruments gate signals from
sources Time (internal), IMMediate (internal), and INPut[1]/INPut2
(Chan 1 and Chan 2 softkeys) can be routed to the rear panel ‘Gate In/ Out’ con-
nector and assigned a polarity using the commands:
OUTPut[:STATe] {OFF|ON}
OUTPut[:STATe]? (query form)
OUTPut:POLarity {NORMal | INVerted}
OUTPut:POLarity? (query form)
ON enables “Gate Out” - gate signals are routed to the rear panel BNC. OFF dis-
ables “Gate Out” - the BNC is an external (“Gate In”) source. Therefore,
OUTPut:STATe OFF must be set when the gate source is EXTernal.
NORMal - the polarity of the gate signal output from the "Gate Out" BNC is a
rising (positive) edge. INVerted - the gate signal polarity is a falling (nega-
tive) edge.