User`s guide

144 53220A/53230A User’s Guide
5 Triggering and Gating
Settings Summary
Table 5-1 is a summary of power-on/reset settings for the trigger and gating
parameters covered in this section.
Table 5-1. Trigger and Gate Settings Summary.
Parameter
Setting Parameter Setting
Trigger
Gate
Trigger source
Trigger slope
Trigger delay
Trigger count
Sample count
Immediate
Negative
0.0s
1
1
Gate start source
Gate external source
Gate start slope
Gate start delay source
Gate start delay events
Gate start delay time
Gate stop source
Gate stop slope
Gate stop holdoff source
Gate stop holdoff events
Gate stop holdoff time
External
External
Negative
Immediate
1
0.0s
External
Positive
Immediate
1
0.0s
Frequency
Frequency gate source
Frequence gate time
Frequency gate polarity
Frequency mode
Frequency burst gate auto
Frequency burst gate delay
Frequency burst gate narrow
Frequency burst gate time
Time
0.1s
Negative
Auto
On
0.0s
Off
1 us
Time Interval
Time interval gate source
Time interval gate polarity
Immediate
Negative
Totalize
Totalize gate source
Totalize gate time
Totalize gate polarity
Time
0.1s
Negative