User`s guide
114 53220A/53230A User’s Guide
3 53220A/53230A Measurements
Carrier Frequency Measurement Examples
// measure the carrier frequency using MEASure? and its
// default settings
*RST // reset to start from known state
MEAS:FREQ:BURS? (@3)
-----------------------------------------------------
// measure the carrier frequency using CONFigure, set
// automatic gate configuration
*RST // reset to start from known state
CONF:FREQ:BURS (@3)
INP3:BURS:LEV -6 // set a detector threshold of -6 dB
SENS:FREQ:BURS:GATE:NARR OFF // narrow pulse mode off
SENS:FREQ:BURS:GATE:AUTO ON // automatic gate set up
READ?
Notes
1. See “Burst Measurement Detector Threshold” in Chapter 4 for information
on setting the threshold for pulse measurements.
2. See “Burst Carrier Frequency Measurements” in Chapter 5 for information
on narrow pulse mode and on automatic and manual gate control for carrier
measurements.
Pulse PRI and PRF
The commands used to measure the pulse repetition interval (PRI) and pulse
repetition frequency (PRF) (Figure 3-13) of a burst signal are: