Specifications

Table Of Contents
Chapter 2 Service
About the Agilent 53210A/53220A/53230A Calibration Menu
98 Assembly-Level Service Guide
NOTE The Auto-Cal command is a subset of the routines executed in
Self-Test (*TST). It is recommended that a counter Self-Test be
performed before running any of the calibration routines. If the
Self-Test passes, the calibration(s) may continue. If the counter
fails the Self-Test, calibration should not be attempted until the
counter is repaired and passes Self-Test. Refer to
“Troubleshooting the Counter” on page 150 .
NOTE It is not necessary to unlock the counter to perform Auto-Cal.
However, any signals routed to the Channel input connectors
should be turned off or disconnected. Also, see the CAUTION
below.
CAUTION A successful ACAL will result in the automatic update of the
compensation factors that a previous ACAL has measured and
stored. Since ACAL changes these compensation factors, if there
is a case where the user is making a measurement (or a series of
sequential measurements) and doesn’t want these compensation
factors changed, the user may not want to run ACAL or Self Test.
To Get Help with the Calibration Menu Functions
1 Press and hold any one of the calibration softkeys.
2 A message will be displayed with context-sensitive Help
information.
Calibrate help is available for the following:
how to calibrate
how to secure the calibration routines
how to change the security code
how to enter the Cal String
NOTE See Figure 2-3 for a summary of the front panel calibration
procedures.