Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Performance Test Record (Tests 1 to 7)
Assembly-Level Service Guide 81
3
Mean
(53220A)
a
b
c
d
Single Period
Test
Ch 1 - 1 MHz
Ch 1 - 250 MHz
Ch 2 - 1 MHz
Ch 2 - 250 MHz
0.999 800 000 uSec
0.003 800 000 uSec
0.999 800 000 uSec
0.003 800 000 uSec
Record: Mean
(53220A)
a. __________________
b. __________________
c. __________________
d. __________________
1.000 200 000 uSec
0.004 200 000 uSec
1.000 200 000 uSec
0.004 200 000 uSec
3
Std Dev
(53220A)
a
b
c
d
Single Period
Test
Ch 1 - 1 MHz
Ch 1 - 250 MHz
Ch 2 - 1 MHz
Ch 2 - 250 MHz
Record: Std Dev
(53220A)
a. __________________
b. __________________
c. __________________
d. __________________
.140 000 000 nSec
.140 000 000 nSec
.140 000 000 nSec
.140 000 000 nSec
3
Mean
(53230A)
a
b
c
d
Single Period
Test
Ch 1 - 1 MHz
Ch 1 - 250 MHz
Ch 2 - 1 MHz
Ch 2 - 250 MHz
0.999 900 000 uSec
0.003 900 000 uSec
0.999 900 000 uSec
0.003 900 000 uSec
Record: Mean
(53230A)
a. __________________
b. __________________
c. __________________
d. __________________
1.000 100 000 uSec
0.004 100 000 uSec
1.000 100 000 uSec
0.004 100 000 uSec
3
Std Dev
(53230A)
a
b
c
d
Single Period
Test
Ch 1 - 1 MHz
Ch 1 - 250 MHz
Ch 2 - 1 MHz
Ch 2 - 250 MHz
Record: Std Dev
(53230A)
a. __________________
b. __________________
c. __________________
d. ____
_______
_______
.028 000 000 nSec
.028 000 000 nSec
.028 000 000 nSec
.028 000 000 nSec
Complete Performance Tests
Test
Number Test
Description
Minimum
Actual Reading
Maximum