Specifications

Table Of Contents
Chapter 1 Performance Tests
Agilent 53210A/53220A/53230A Complete Performance Tests
76 Assembly-Level Service Guide
4 Record the Mean and Standard Deviation for Test 7c, in the
“Agilent 53210A/53220A/53230A Performance Test Record (Tests 1
to 7)” starting on page 78?.
5Press the Reset Stats softkey.
NOTE For Quick Verification (Q), only the narrow pulse width burst
frequency test needs to be performed.
6 Set the E8257D-520 to a pulse width of 13 uSec (Wide Pulse).
7Press [Freq/Period] / (Pulse/Burst) / (Burst Freq) keys.
8 Press and set:
•dB Level: -6
Narrow: Off
•Gate Auto: On
9Press the Trigger key.
The counter takes 100 readings and fills in the math functions.
10 Verify the ‘Mean’ is:
1.35 GHz ± 20.8 kHz
11 Verify the ‘Std Dev’ is less than:
•7.79 kHz
12 Record the Mean and Standard Deviation for Test 7d, in the
“Agilent 53210A/53220A/53230A Performance Test Record (Tests 1
to 7)” starting on page 78.
13 Press the Reset Stats softkey.
14 Disconnect the test setup.