14 Press to set:
Statistics: On
The display shows:
Channel 3 Pulse Burst Frequency Test Procedure
1Press the Trigger key.
The counter takes 100 readings and fills in the math functions.
2 Verify the ‘Mean’ is:
1.35 GHz ± 1.35 MHz
3 Verify the ‘Std Dev’ is less than:
338 kHz